Product Details:
|
Product Category: | Safety Testing System | Warranty: | 12 Months |
---|---|---|---|
OS: | Windows XP | Color: | Grey (customized) |
Power Supply: | Customized As The National Grid | Monitor: | 5.7” LCD Blue Background |
High Light: | electrical safety test equipment,laboratory testing equipment |
Safety Testing System / Microwave Oven
1. Test object: microwave oven
2. Technical specification
Test items |
AC Hipot, DC Hipot, Insulation resistance, Grounding resistance, leakage, power |
Data save mode |
USB save |
Insulation resistance |
|
Output voltage setting range/ precision |
DC 500/1000V ±(3%×reading +5 words) |
Insulation resistance testing range / precision |
1~2000MΩ ≤100MΩ:±(5%×display value +3 words) >100 MΩ:±(8%×display value +8words) |
Insulation resistance alarm setting range |
Max:0~2000MΩ;Min:1~2000MΩ |
Test time range/ resolution |
1~300s 1s/step |
AC Hipot |
|
Output voltage setting range/ precision |
AC 200~5000V ±(3%×setting value+5 words) |
Breakdown current testing range/ precision |
0.10~25.00mA ±(3%×display value+5 words) |
Breakdown current presetting alarm range |
Max:0.10~25.00mA ;Min:0.00~5.00mA |
Test time range/ resolution |
1~300s 1s/step |
Leakage |
|
Output voltage |
1.06 times rated voltage |
Leakage type setting |
Dynamic leakage, static leakage, optional equipped with 3 types of body network, which can be customized and switched freely. |
Output voltage testing range / precision |
AC 50~280V ±(0.5%×display value+2words) |
Leakage current test range/ precision |
0.05~20.00mA ±(1.5%×display value +0.02mA) |
Leakage current setting alarm range |
Max:0.05~20.00mA;Min:0.01~5.00mA |
Test time range |
1~300s 1s/step |
Power |
|
Output voltage |
1.0 times rated voltage |
Output voltage testing range / precision |
AC 50~280V ±(0.5%×display value+2words) |
Output current testing range / precision |
AC 0.20~30.00A ±(0.5%×display value+2words) |
Active power testing range / precision |
30~6000W ±(0.5%×display value+5words) |
Multi-position setting |
4 positions can be set, identify position switch automatically, each position can display corresponding test data after complete the test. |
Power setting alarm range |
Max:3~6000W Min:0~6000W |
Test time range / resolution |
1~300s 1s/step |
Grounding resistance |
|
Grounding current output range/ precision |
AC 5.0~30.0A ±(3%×setting value +5 words) |
Grounding resistance testing range/ precision |
10.0~600mΩ ±(3%×display value+5 words) |
Grounding resistance alarm setting range |
Max:5~10A:10.0~600mΩ;11~25A:10.0~300mΩ;26~30A:10.0~200mΩ; Min:0.0~100mΩ |
Test time range/ resolution |
1~60s 1s/step |